Niscomed M1000 Multipara Patient Monitor Firmware

CPE Details

Niscomed M1000 Multipara Patient Monitor Firmware
-
2020-10-08
13h31 +00:00
2020-10-08
13h31 +00:00
Alerte pour un CPE
Stay informed of any changes for a specific CPE.
Notifications manage

CPE Name: cpe:2.3:o:niscomed:m1000_multipara_patient_monitor_firmware:-:*:*:*:*:*:*:*

Informations

Vendor

niscomed

Product

m1000_multipara_patient_monitor_firmware

Version

-

Related CVE

Open and find in CVE List

CVE ID Published Description Score Severity
CVE-2020-15485 2020-08-26 17h12 +00:00 An issue was discovered on Nescomed Multipara Monitor M1000 devices. The onboard Flash memory stores data in cleartext, without integrity protection against tampering.
5.5
Medium
CVE-2020-15482 2020-08-26 13h44 +00:00 An issue was discovered on Nescomed Multipara Monitor M1000 devices. The device enables an unencrypted TELNET service by default, with a blank password for the admin account. This allows an attacker to gain root access to the device over the local network.
7.8
High
CVE-2020-15483 2020-08-26 13h37 +00:00 An issue was discovered on Nescomed Multipara Monitor M1000 devices. The physical UART debug port provides a shell, without requiring a password, with complete access.
6.8
Medium
CVE-2020-15484 2020-08-26 12h41 +00:00 An issue was discovered on Nescomed Multipara Monitor M1000 devices. The internal storage of the underlying Linux system stores data in cleartext, without integrity protection against tampering.
7.5
High